HUBEI AGRICULTURAL SCIENCES ›› 2019, Vol. 58 ›› Issue (13): 111-113.doi: 10.14088/j.cnki.issn0439-8114.2019.13.023

• Detection Analysis • Previous Articles     Next Articles

An improved method for calibrating the concentration of Na2S2O3 solution

LI Wei, XIONG Jian, ZHOU Yi, YAN Can   

  1. Science School,Tibet University,Lhasa 850000,China
  • Received:2018-12-24 Online:2019-07-10 Published:2019-12-06

Abstract: In order to study the optimum conditions for calibrating the concentration of Na2S2O3 solution with known concentration of H2O2 solution, the results were analyzed by single factor and orthogonal experiment. The experimental results showed that the concentration of 0.821 8 mol/L H2O2 solution 1.00 mL,20 g/L KI solution 60 mL and 25%H3PO4 6 mL solution for 30 min. The experimental results were the best, the standard deviation s was 0.09%, and compared with the results of K2Cr2O7 method. The experimental results were analyzed by F test, no significant difference. The method of green environmental protection, can be used as an method of calibrating Na2S2O3 solution concentration.

Key words: Na2S2O3 solution, calibration, method, improvement